Trying Out the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
58Getting Started 1330.8073.02 ─ 03
b) Tap the "Frequency" button.
c) In the "Center" field, enter 64 on the number pad on the front panel.
d) Press the "MHz" key next to the number pad.
2. Reduce the span to 20 MHz:
a) In the "Span" field of the "Frequency" dialog box, enter 20
MHz.
b) Close the "Frequency" dialog box.
3. Set the reference level to -25 dBm:
a) In the configuration "Overview", tap the "Amplitude" button.
b) In the "Value" field of the "Amplitude" dialog box, enter -25
dBm.
The display of the calibration signal is now improved. The maximum at the
center frequency (=calibration frequency) of 64 MHz becomes visible.
Figure 6-2: Calibration signal with optimized display settings
6.2 Displaying a Spectrogram
In addition to the standard "level versus frequency" spectrum display, the
R&S FSV/A also provides a spectrogram display of the measured data. A spec-
trogram shows how the spectral density of a signal varies over time. The x-axis