Trying Out the Instrument
R&S
®
FSVA3000/R&S
®
FSV3000
57Getting Started 1330.8073.02 ─ 03
5. Tap the "Calibration Frequency RF" option. Leave the frequency at the default
64 MHz, with a narrowband spectrum.
The calibration signal is now sent to the RF input of the R&S FSV/A. By
default, a continuous frequency sweep is performed, so that the spectrum of
the calibration signal is now displayed in the standard level versus frequency
diagram.
Figure 6-1: Calibration signal as RF input
Instrument warmup time
Note that the instrument requires an initial warmup time after switching it
on. A message in the status bar ("Instrument warming up...") indicates that
the operating temperature has not yet been reached. Wait until this mes-
sage is no longer displayed before you start a measurement.
To optimize the display
To optimize the display for the calibration signal, we will adjust the main measure-
ment settings.
1. Set the center frequency to the calibration frequency:
a) Tap the "Overview" softkey to display the configuration "Overview".
Measuring a Basic Signal