R&S NRP-Z11/21/31/41/61/211/221 Remote Control - Commands
1137.7470.12 6.5 E-7
SENSe (Sensor Configuration)
The sensor is configured by means of the commands of the groups SENSe and TRIGger.
Table 6-2 Commands of the SENSe system
Command Parameter Unit Remarks
SENSe
:AVERage
:COUNt[?] 1 to 65536
:AUTO[?] OFF | ON | ONCE
:MTIMe[?] 0.01 to 999.99 s
:NSRatio[?] 0.0 to 1.0 dB
:RESolution[?] 1 to 4
:SLOT[?] 1 to <SENSe:POWer:TSLot:AVG:COUNt>
:TYPE[?] RESolution | NSRatio
:RESet No query
:STATe[?] OFF | ON
:TCONtrol[?] MOVing | REPeat
:CORRection
:DCYCle[?] 0.001 to 99.999 %
:STATe[?] OFF to ON
:OFFSet[?]200.0 to 200.0 dB
:STATe[?] OFF to ON
:SPDevice:STATe[?] OFF to ON
:FREQuency[?] 10.0e6 to 8.0e9 (NRP-Z11/211) or
10.0e6 to 18.0e9 (NRP-Z21/221) or
10.0e6 to 33.0e9 (NRP-Z31) or
10.0e6 to 40.0e9 (NRP-Z41) or
10.0e6 to 50.0e9 (NRP-Z61)
Hz
:FUNCtion[?] POWer:AVG“ | “POWer:TSLot:AVG“ |
“POWer:BURSt:AVG“ | “XTIME:POWer“
:POWer
:AVG
:APERture[?] 10.0e-6 to 0.3 s