Remote Control - Commands R&S NRP-Z11/21/31/41/61/211/221
1137.7470.12 6.14 E-7
SENSe:POWer:BURSt:DTOLerance[?] 0.0 to 0.003
SENSe:POWer:BURSt:DTOLerance defines the dropout tolerance, a parameter for reliably detecting
the burst end in the Burst Average mode with modulated signals (e. g. with digital standards NADC,
PDC, PHS, etc.). The dropout tolerance should be selected larger than the longest amplitude drop and
smaller than the gap between two consecutive bursts. The default value is sufficient for all usual digital
communication standards.
The query yields the dropout tolerance for the Burst Average mode.
Default setting: 0.0001 [s]
Dropout time
Trigger
level
Power
Time
BURSt:DTOLerance
Pulse duration
BURSt:DTOLerance
Fig. 6-1 Effect of SENSe:POWer:BURSt:DTOLerance
SENSe:POWer:TSLot:AVG:COUNt[?] 1 to 128
For the Timeslot mode, SENSe:POWer:TSLot:AVG:COUNt sets the number of consecutive timeslots
that are to be processed after each trigger event.
The query yields the number of consecutive timeslots.
Default setting: 8
SENSe:POWer:TSLot:AVG:WIDTh[?] 10.0e–6 to 0.1
SENSe:POWer:TSLot:AVG:WIDTh sets the length of a timeslot for the Timeslot Average mode (in s).
The query yields the length of a timeslot for the Timeslot Average mode.
Default setting: 0.001 [s]