R&S NRP-Z11/21/31/41/61/211/221 Remote Control - Commands
1137.7470.12 6.23 E-7
Item Information string Remarks
"CAL. S-PARA." "<date>" Date of s-parameter calibration in the format YYYY-MM-DD. If
no S parameter set is loaded, the sensor returns the string "not
applicable". "Invalid Calibration Date" is returned with an invalid
date entry.
"CAL. MISC." "<date>" Date of the calibration of other parameters in the format YYYY-
MM-DD. "Invalid Calibration Date" is returned with an invalid
date entry.
"CAL. TEMP." "not applicable" Temperature response calibration of the R&S NRP-
Z11/21/31/41/61/211/221 is neither required nor possible.
"CAL. LIN." "not applicable" Linearity calibration of the R&S NRP-Z11/21/31/41/61/211/221
is neither required nor possible.
"SPD MNEMONIC" "<mnemonic string>" Clear-text designation of the components connected ahead of
the sensor.
"TESTLIMIT" "0.160 dB" Test limit for the automated test conducted by the NRP power
meter base unit if s-parameter correction has been deactivated.
"TESTLIMIT PD." "<testlimit>" Test limit for the automated test conducted by the NRP power
meter base unit if s-parameter correction has been activated
(actual value dependent on s-parameters of the component
connected ahead of the sensor).
"UPTIME" "<uptime in s>" Uptime of the power sensor since the last power-up/firmware
update in seconds.
SYSTem:INITialize
SYSTem:INITialize sets the sensor to the standard state, i.e. the default settings for all test parameters
are loaded in the same way as with *RST. The sensor then outputs a complete list of all supported
commands and parameters. With the command, the remote-control software can automatically adapt to
the features of different types of sensors with different functionality.
SYSTem:MINPower?
SYSTem:MINPower? yields the lower test limit of the sensor or the combination comprising the sensor
and components connected ahead of it, if the SENSe:CORRection:SPDevice parameter has the ON
value. This query can be used to determine a useful resolution for the result display near the lower test
limit.
SYSTem:RUTime[?] 0.0 to 10.0
This command is used to limit the output rate of measurements with continuous output of measurement
results (setting INITiate:CONTinuous ON). This is practical if the measurement time is very short or if
intermediate values are output (command SENSe:AVERage:TCONtrol MOVing or
SENSe:TRACe:AVERage:TCONtrol MOVing). Without this limitation, the controlling host can be
overloaded very fast or excessively occupied with accepting measured values.