R&S NRP-Z11/21/31/41/61/211/221 Remote Control - Commands
1137.7470.12 6.17 E-7
The query yields
• 1 for OFF,
• 2 for ON.
Default setting: OFF
SENSe:SGAMma:MAGNitude[?] 0.0 to 1.0
SENSe:SGAMma:MAGNitude defines the magnitude of the complex reflection coefficient of the signal
source. A value of 0.0 corresponds to ideal match and a value of 1.0 to total reflection.
The query yields the set magnitude.
Default setting: 0.0
SENSe:SGAMma:PHASe[?] –360.0 to 360.0
SENSe:SGAMma:MAGNitude defines the phase angle (in degrees) of the complex reflection coefficient
of the signal source.
The query yields the set phase angle.
Default setting: 0.0 [°]
SENSe:TIMing:EXCLude:STARt[?] 0.0 to 0.1
SENSe:TIMing:EXClude:STARt defines the exclusion time at the beginning of the measurement
window in the Burst Average and Timeslot Average modes. It is indicated by T
START
in Figs. Fig. 6-4 and
Fig. 6-5.
The query yields the exclusion time at the beginning of the measurement window.
Default setting: 0.0 [s]
SENSe:TIMing:EXCLude:STOP[?] 0.0 to 0.003
SENSe:TIMing:EXClude:STOP defines the exclusion time at the end of the measurement window in the
Burst Average and Timeslot Average modes. It is indicated by T
STOP
in Figs. Fig. 6-4 and Fig. 6-5.
The query yields the exclusion time at the end of the measurement window.
Default setting: 0.0 [s]