R&S NRP-Z11/21/31/41/61/211/221 Remote Control - Commands
1137.7470.12 6.7 E-7
Command Parameter Unit Remarks
:REALtime[?] OFF | ON
:TIME[?] 0.0001 to 0.3
SENSe:AVERage:COUNt[?] 1 to 65536
SENSe:AVERage:COUNt sets the number of measured values that have to be averaged for forming
the measurement result in the modes Continuous Average, Burst Average and Timeslot Average. The
higher this averaging factor, the less the measured values fluctuate and the longer the measurement
time lasts. The parameter is rounded off to the nearest power-of-two number.
The query yields the averaging factor used in the modes Continuous Average, Burst Average and
Timeslot Average.
The averaging function must be activated with SENSe:AVERage:STATe ON so that t
set averaging factor becomes effective.
Default setting: 4
SENSe:AVERage:COUNt:AUTO[?] OFF | ON | ONCE
SENSe:AVERage:COUNt:AUTO activates (auto-averaging) or deactivates (fixed-filter mode) automatic
determination of the averaging factor in the modes Continuous Average, Burst Average and Timeslot
Average. If auto-averaging is activated, the averaging factor is continuously determined and set
depending on the level of power and other parameters.
SENSe:AVERage:COUNt:AUTO ON activates auto-averaging and SENSe:AVERage:COUNt
:AUTO OFF deactivates it. On deactivation, the previous, automatically determined averaging factor is
used in the fixed-filter mode. The SENSe:AVERage:COUNt:AUTO ONCE command ensures that a new
averaging factor is determined by the filter automatic function under the current measurement
conditions and used in the fixed-filter mode.
The query yields
• 1 for OFF,
• 2 for ON.
Default setting: ON
SENSe:AVERage:COUNt:AUTO:MTIMe[?] 0.01 to 999.99
SENSe:AVERage:COUNt:AUTO:MTIMe sets an upper limit for the settling time of the auto-averaging
filter in the NSRatio mode and thus limits the length of the filter.
The query returns the time that has been set.
Default setting: 4.0 [s]