Remote Control - Commands R&S NRP-Z11/21/31/41/61/211/221
1137.7470.12 6.30 E-7
The query yields
1 for POSitive,
2 for NEGative.
Default setting: POSitive
TRIGger:SOURce[?] HOLD | IMMediate | INTernal | BUS | EXTernal |
TRIGger:SOURce sets the trigger source.
HOLD: Triggering only with command TRIGger:IMMediate.
IMMediate: Automatic triggering without explicit event.
INTernal: Triggering by the measurement signal. Relevant parameters:
TRIGger:LEVel, TRIGger:SLOPe, TRIGger:DELay, TRIGger:HYSTeresis,
TRIGger:HOLDoff.
BUS: Triggering with command *TRG or TRIGger:IMMediate.
EXTernal: Triggering via the hardware trigger bus, e.g. via the R&S NRP-Z5 sensor hub
or the R&S NRP2 base unit. Relevant parameters:
TRIGger:SLOPe, TRIGger:DELay, TRIGger:HOLDoff.
The query yields
1 for HOLD,
2 for IMMediate,
4 for INTernal,
8 for BUS,
16 for EXTernal.
Default setting: IMMediate