Test Procedures R&S SMU200A
1007.9845.82 1.74
E-12
Adjacent Channel Power for 3GPP FDD
The equipment layout for 3GPP FDD signal generation includes the options R&S B13 (Baseband Main
Module), R&S B10 (Baseband Generator) and R&S K42 (Digital Standard 3GPP FDD).
Test equipment Signal analyzer R&S FSQ including option R&S FSQ-K70
(Vector Signal Analysis) (table 1-1, pos. 24)
Test setup
Connect signal analyzer to RF output of DUT.
Test method Set standard 3GPP FDD in the baseband. Perform the measurements
with the 3GPP measurement setting of the R&S FSQ.
Measurement Settings on DUT
Level: max. level, refer to data sheet (depending on options)
Frequencies: 2160 MHz (and 5 GHz for R&S B106)
3GPP FDD:
Test Setups Test_Model_1_64channels
State ON
I/Q Settings
Manual Gain Unicod 3GPP ACP optimized 3 dB
Settings on analyzer
FREQ CENTER: 2160 MHz (and 5 GHz for R&S B106)
MEAS CHAN PWR ACP
CP /ACP STANDARD WCDMA 3GPP FWD
ADJUST REF LVL
NOISE CORR ON
SWEEP SWEPP TIME MANUAL 1 sec
Check ACP
:
A
djacent Channel (take the smaller of the two measurement values
UPPER/LOWER)
Settings on DUT
I/Q Settings
Manual Gain Unicod Low Noise 6 dB
Check ACP
:
A
lternate Channel (take the smaller of the two measurement values
UPPER/LOWER)
The number of necessary measurement depends on the number of installed RF- and baseband paths,
see following table:
Installed paths
1 x RF
1 x BB
2 x RF
2 x BB
2 x RF
1 x BB
1 x RF
2 x BB
Modulation error 3 GPP FDD
Number of measurements
1x 2x 2 x 2 x