Installing the Options R&S SMU200A
1007.9845.32 4.6 E-3
3. Select the Select Backup window with the rotary knob.
4. Select the backup to be deleted with the up/down cursor keys and the ENTER key.
5. Select the Remove button with the rotary knob
6. Start the deletion with the ENTER key
After the deletion, the instrument returns to the Remove Backup menu as long as backups are still
available. If the last backup is deleted the Windows XP Recovery and Backup Partition menu
opens again.
7. Select the Cancel button with the rotary knob
8. Close the Remove Backup menu with the
ENTER key.
The Windows XP Recovery and Backup Partition menu opens.
9. Select the Exit and Shutdown button with the rotary knob.
10. Exit the menuand shutdown the instrument with the ENTER key.
Installing the Options
A list of all available R&S SMU options is provided in the data sheet and on the internet
(www.smu200a.rohde-schwarz.com)
.
Hardware Options
Installation and replacement of hardware options is described in chapter 3 of this service manual.
Please also note the mounting instructions enclosed with the options. These mounting instructions can
be filed at this place in the service manual and are thus easily available whenever they are required.
CAUTION
Disconnect the instrument from the mains before opening the casing. Also note the
safety instructions at the beginning of this manual.
The components used in the instrument are sensitive to electrostatic discharges
which is why they are to be dealt with according to the ESD regulations.
The SETUP Installed Options menu provides information on the already installed options.