R&S SMU200A Troubleshooting
1007.9845.82 3.39 E-6
Error messages concerning the module IQOP3
Error Error correction
"Output unleveled, RMS detector cold"
The temperature control of the detector is unable to
maintain the desired temperature of 70 °C. RF Level is out
of specification. This can be checked using diagnostic
system:
Thermostat temperature is displayed using test point
SETUP - Test Point DIAG_IQOP3_RFM_TEMP_DET
100 * U
DIAG
= Thermostat temperature in degrees Celsius
If temperature deviates more than 4 °, the module is to be
replaced.
"Output unleveled: Gain Control on Upper
Limit"
"Output unleveled: Gain Control on Lower
Limit"
"Output unleveled: Gain out of Setting
Range"
Check settings, especially of attenuator mode and ALC
mode. Start anew with blue PRESET key.
Try to execute internal adjustmentAdjust Level. See
section “Troubleshooting – Internal Adjustments”.
Check for correct input levels from SSYN or IQOP6.
"Questionable level adjustment data"
"Questionable IQOP3 I/Q adjustment
data"
"Questionable IQOP6 I/Q adjustment
data"
"Baseband adjustment failed"
"Zero adjustment failed"
"AM adjustment failed"
"IQOP3 I/Q adjustment failed"
"IQOP6 I/Q adjustment failed"
These error messages indicate problems with internal
adjustment prodedures. Retry internal adjustments (see
section “Troubleshooting – Internal Adjustments”).
If error still exists, check for correct input signals (see
above).
"Missing level adjustment data"
"Missing IQOP3 I/Q adjustment data"
"Missing IQOP6 I/Q adjustment data"
There are no data of internal adjustments. This may be the
result of problems with the hard disk. Execute internal
adjustments (see sectionTroubleshooting Internal
Adjustments”).
"Impairments DAC on limit"
The desired setting of impairments is limited. If this
happens with no impairment set, execute internal
adjustmentAdjust I/Q Modulator (see section
“Troubleshooting – Internal Adjustments”).
"Vector Modulator: Quadrature Loop
unlocked"
The submodule RFM is defective and must be changed.