R&S SMU200A Test Procedures
1007.9845.82 1.57
E-12
Baseband Digital IQ-Output (Option R&S -B18)
Test equipment
Signal generator (Table 1-2, item 5), equiped with Digital Input
(B17)
Cable for Digital-Baseband IO (Table 1-2, item 30),
Test setup
Connect signal generator’s Digital Input to the Digital Output of
the DUT with the Mini D Ribbon cable.
Test method
Set the Digital Output of the DUT in test sequence t
ransmitting mode.
The signal generator's Digital Input evaluates the incoming test pattern
and calculates the Bit Error Rate.
Measurement Settings on DUT
Activate test sequence transmitting via remote control with
command :TEST<HW>:BBOut:TTEST ON
Settings on signal generator
Start bit error evaluation via remote control with command:
:TEST<HW>:BBIN:RBERror?
Result:
0 = TestOk,
1 = TestFailed
Test will take 2 seconds to deliver a result. Repeat it 3 times, all of
the test results must be OK.
Settings on DUT
Deactivate test sequence transmitting via remote control with
command :TEST<HW>:BBOut:TTESt OFF
Remark: <HW> = 1 for path 1, only one B18 can be equiped

Table of Contents