R&S SMU200A Module Replacement
1007.9845.82 3.47 E-6
Note: The words “left” and “right” in the manual always refer to the front view of the instrument.
Safety Instruction, Dismounting the Case
Caution!
Wrist strap with cord
Building ground
Floor mat
Ground connection
of operational site
Heel strap
Please note the safety instructions at the beginning of
this manual.
Disconnect the instrument from the mains before
opening the case.
Protect the replacement site against electrostatic
discharge to avoid damage to electronic components of
the modules.
The following two methods of ESD protection may be used
together or separately:
Wrist strap with cord to ground connection
Conductive floor mat and heel strap combination
Put the instrument on the front handles before loosening
the rear feet and pulling off the case to avoid damage to
the instrument.
When mounting the case, take care not to damage or
pull off cables.
The case is to be dismounted for all the following procedures and retrofitted to finish. This will not be
mentioned further.