R&S SMU200A Baseband Digital IQ-Output (Option R&S SMU/SMJ-B18)
1007.9845.82 1.91
E-12
AUX I/O and BNC Connectors
Test equipment - Adapter R&S SMU-Z5 (AUX I/O to BNC) (table 1-1, Pos. 13);
several BNC cable
Test setup Plug adapter into AUX I/O interface
Make the following BNC connections on the adapter:
PARDATA0 USER2
PARDATA1 USER3
PARDATA2 USER4
PARDATA4 BURST
LEVATT INST_TRIG_B (HOP) on the rear side of the device
Make the following BNC connections on the instrument and the BERT
adapter:
from: to
:
M
ARKER1A TRIGGER1, BER CLK, BER_RESTART
MARKER2A TRIGGER2, BER_DATA, BER_EN_DATA
MARKER1B USER1
Test method Self-test by means of connection test
Measurement Settings on DUT
Custom Digital Mod:
State ON (the markers are available in this state only)
Start selftest AUX-
IO and BNC via remote control with commands
:TEST:CONNector:AUXio and :TEST:CONNector:BNC.
The number of necessary measurement depends on the number of installed RF- and baseband paths,
see following table:
Installed paths
1 x BB 2 x BB
Custom Digital Modulation: trigger and marker function
Number of measurements
1 x 1 x