R&S SMU200A Baseband Digital IQ-Output (Option R&S SMU/SMJ-B18)
1007.9845.82 1.93
E-12
Adjacent Channel Power for 3GPP FDD
The equipment layout for 3GPP FDD signal generation includes the options R&S SMU-B13 (Baseband
Main Module), R&S
SMU-B9/10/11 (Baseband Generator) and R&S SMU-K42 (Digital Standard 3GPP
FDD).
Test equipment
- See section
See section „Test Assembly for Measurement realized
RF Evaluation“ on page 1.2
- Signal analyzer R&S FSQ including option R&S FSQ-K70
(Vector Signal Analysis) (table 1-1, pos. 24)
- Signal generator if no
R&S SMU-B10 (Baseband Generator)
is installed
Test setup
Connect the DUT I/Q-outputs to the I/Q-inputs of the signal
generator
Connect the signal analyzer to the RF output of the signal
Generator
.
Test method Set standard 3GPP FDD in the baseband. Perform the measurements
with the 3GPP measurement setting of the R&S FSQ.
Measurement Settings on signal generator
Level: 0 dBm
Frequency: 2160 MHz
RF: State ON
I/Q Settings menu:
Source Analog Wideband I/Q Input
State On
Settings on DUT
Level: max. level, refer to data sheet (depending on options)
Frequencies: 2160 MHz
3GPP FDD:
Test Setups Test_Model_1_64channels
State ON
I/Q Settings
I/Q Gain 3GPP ACP optimized 3 dB
Settings on analyzer
FREQ CENTER: 2160 MHz
MEAS CHAN PWR ACP
CP /ACP STANDARD WCDMA 3GPP FWD
ADJUST REF LVL
NOISE CORR ON
SWEEP SWEPP TIME MANUAL 1 sec
Check ACP:
Adjacent Channel (take the smaller of the two measurement
values UPPER/LOWER)
Settings on DUT
I/Q Settings
I/Q Gain Low Noise 6 dB
Check ACP:
Alternate Channel (take the smaller of the two measurement
values UPPER/LOWER)