R&S SMU200A Test Procedures
1007.9845.82 1.13
E-12
Nonharmonics
Test equipment Same as for harmonics
Test setup Same as for harmonics
Measurement
Setting on analyzer:
AMPT/MIXER/MIXER LVL MANUAL –5 dBm
First the level is measured at the test frequency as reference, th
en a
s
ignal is searched for at the analyzer frequency.
The nonharmonic spacing is the measured level referred to the
reference level (dBc = referred to the carrier).
Note:
Some of the nonharmonics suppression values to be measured are
outside analyzer specifications. In case of doubt, the measurement
should be repeated with a 3 dB attenuator at the analyzer input. If
the nonharmonic spacing changes, the nonharmonic comes from
the analyzer.
Recommended settings and sampling frequencies:
Nonharmonics of the synthesis without option R&S B22 at test level 3.0 dBm, unmodulated
Setting on DUT
Frequency in MHz
Analyzer search frequency in MHz
984.001 987.557
1007.6443 1007.6515
1027.155 1027.20444
1144.005 1144.055
1253.0 1265.2
1401.26 1401.3074
1499.02379 1499.03469
1696.853 1696.8639
1849.5328 1849.5438
2696.296 2696.3058
Nonharmonics of output mixer without option R&S B22, test level P
max
, unmodulated
Setting on DUT
Frequency in MHz
Analyzer search frequency in MHz
200.0 1240.00
200.0 640.00
200.0 440.00
200.0 240.00