R&S SMU200A Test Procedures
1007.9845.82 1.29
E-12
Frequency response
Test equipment AC voltmeter (Table 1-2, item 19)
Test setup
Connect the AC voltmeter to the LF socket of the DUT.
Measurement
Settings on DUT
:
LF
Output menu:
Vary LF Gen Frequency from 10 Hz to 1 MHz.
Recommendation: logarithmic with 4 steps/decade.
Measure the output level
.
T
he frequency response is the difference between the highest and
the lowest level.
Note: The settling time is a pure computer time and need therefore not be measured.