R&S SMU200A Test Procedures
1007.9845.82 1.53
E-12
Adjacent Channel Power for 3GPP FDD
The equipment layout for 3GPP FDD signal generation includes the options R&S B13 (Baseband Main
Module), R&S B10 (Baseband Generator) and R&S K42 (Digital Standard 3GPP FDD).
Test equipment Signal analyzer R&S FSQ including option R&S FSQ-K70 (Vector Signal
Analysis) (table 1-1, pos. 24)
Signal generator (Table 1-1, item 5)
Test setup
Connect signal generator’s I/Q-outputs to I/Q-inputs of DUT
.
C
onnect signal analyzer to RF output of DUT.
Test method Set standard 3GPP FDD in the baseband of the signal generator and feed
the I/Q signal into the DUT’s I/Q-inputs. Perform the measurements with
the 3GPP measurement setting of the R&S FSQ.
Measurement
Settings on DUT
Level: max. level, refer to data sheet (depending on options)
Frequencies: 2160 MHz (and 5 GHz for R&S B106)
I/Q Settings menu:
Source Internal Bas
eband
S
tate On
BB In Settings menu:
State On
Mode analog i
nput
C
rest Factor 0 dB
Peak Level 0 dBfs
I/Q Setting
s
M
anual Gain 3GPP ACP optimized 3 dB
Settings on signal generator
3GPP FDD:
Test Setups Test_Model_1_64channels
State ON
Settings on analyzer
FREQ CENTER: 2160 MHz (and 5 GHz for R&S B106)
MEAS Z CHAN PWR ACP
Z CP /ACP STANDARD WCDMA 3GPP FWD
Z ADJUST REF LVL
Z NOISE CORR ON
SWEEP Z SWEPP TIME MANUAL 1 sec
Check ACP
:
A
djacent Channel (take the smaller of the two measurement
values UPPER/LOWER)
Settings on DUT
I/Q Settings
Manual Gain Low Noise 6 dB
Check ACP
:
A
lternate Channel (take the smaller of the two measurement
values UPPER/LOWER)
Test results should not miss the typical values by more than 4 dB.
The measurement has to be carried out for each RF path installed in the DUT.