R&S SMU200A Test Procedures
1007.9845.82 1.55
E-12
Output voltage
Test equipment AC/DC Voltmeter R&S URE 3 (table 1-1, pos. 19).
Test setup
Connect the voltmeter to the I/(and Q)-output of the DUT
.
Test method
A single sideband signal is generated using the Custom Digital
Modulation functionality by selecting only one carrier (DC-Signal), the
offset is set with the Frequency Offset function in the Baseband menu.
The test levels are set one after the other.
Measurement Settings on DUT
Custom Digital Modulation: State ON
Data Source: All 0
Modulation Type:
BPSK
F
ilter: Rectangular
Frequency Offset: 100 kHz
I/Q Out Menu:
Output Type: Single Ended
Mode: Variable
I/Q I Level EMF: 1 V
Settings on multimeter
COUPLING A
C
D
ETEKTOR RM
S
M
ultiply measured voltage with [2.
Measurement level: 2 V (reference), 1 V, 0.4 V and 0.2 V
.
Bias voltage
Test equipment AC/DC Voltmeter R&S URE 3 (table 1-1, pos. 19).
Test setup
Connect the multimeter to the I/(and Q)-output of the DUT
.
Test method The bias voltage is measured for different values.
Measurement Settings on DUT
I/Q Out Menu:
Output Type: Single Ended
Mode: Variable
I/Q I Level EMF: 0.1 V
Couple I/Q Bias: On
I Bias EMF: measurement bias
Settings on multimeter
COUPLING DC
Measurement Check bias: -3.6, -1, -0.3, -0.1, -0.03, -0.01, 0, +0.002, 1
and 3.6 V