R&S SMU200A Test Procedures
1007.9845.82 1.65
E-12
Function External Clock
The equipment layout for generating the digital modulation signals includes the options R&S B13 (Baseband
Main Module) and R&S B10 (Baseband Generator).
Test equipment Vector Signal Generator DUT (table 1-1, pos. Pos. 5) ,
Signal Analyzer R&S FSQ with option R&S FSQ-K70
(vector signal analysis).
Test setup
Connect CLOCK outputs of reference DUT to CLOCK inputs of tes
t
DU
T (DUT)
.
Connect signal analyzer to RF output of DUT
Test method The test DUT is supplied with a clock from the reference DUT via the
external interface. The modulated signal is extracted at the RF output of
the DUT, demodulated with the R&S FSQ and checked for EVM.
Measurement Settings on DUT
Frequency: 100 MHz
Level: 0 dBm
Custom Digital Mod:
State ON
Data Source Pattern 010101010101
Modulation Type 8PSK
Symbol Rate 20 MHz
Filter RRC (Root-Raised-Cosine)
Roll off 0.22
Clock
Source: External
Mode: Multiple Symbol
Multiplier: 5
Global Settings
Threshold Data Input 50 Ohm
Impedance Data Input 1.5 V
Settings on reference-DUT
Frequency: 100 MHz
Level: 0 dBm
ARB
Load Waveform any
State ON
Clock Freq 100 MHz
Check external clock rate (Measured Extern Clock in Trigger-
Marker-Clock menu) - must be 100 MHz.
Settings on analyzer:
VSA
FREQ CENTER: 100 MHz
AMPT / REF LEVEL 5 dBm
MODULATION SETTINGS
SYM RATE 20 MHz
MODULATION & MAPPING PSK 8PSK
MODULATION FILTER RRC RRC RRC RRC
ALPHA BT 0.22
Check EVM in Modulation Accuracy Tabl
e.
The number of necessary measurement depends on the number of installed RF- and baseband paths, see
following table:
Installed paths
1 x RF
1 x BB
2 x RF
2 x BB
2 x RF
1 x BB
1 x RF
2 x BB
Custom Digital Modulation: external clock function
Number of measurements
1 x 1 x 1 x 1 x