Test Procedures R&S SMU200A
1007.9845.82 1.70
E-12
Measurement 2 Settings on DUT:
Level: max. level, refer to data sheet (depending on options)
Frequencies: 910 MHz, 1850 MHz
(and 5 GHz for R&S B106)
GSM/EDGE:
State ON
Settings on analyzer
GSM/EDGE
FREQ CENTER 910 MHz, 1850 MHz (and 5 GHz)
GSM/EDGE
DEMOD SETTINGS
AUTO LEVEL&TIME
GSM/EDGE
POWER VS TIME
START REF MEAS.
LIMIT CHECK PASSED has to be indicated on the analyzer.
Settings on analyzer
GSM/EDGE
TRANSIENT SPECTRUM
START REF MEAS
LIMIT CHECK PASSED has to be indicated on the analyzer.
Settings on analyzer
GSM/EDGE
MODULATION SPECTRUM
START REF MEAS
LIMIT CHECK PASSED or MARG has to be indicated on the analyzer.
The number of necessary measurement depends on the number of installed RF- and baseband paths, see
following table:
Installed paths
1 x RF
1 x BB
2 x RF
2 x BB
2 x RF
1 x BB
1 x RF
2 x BB
GSM and GSM Normal Burst
Number of measurements
1x 2x 2 x 2 x