55
The Density Calibration is complete.
For future reference, document these items:
(a) Environmental/process conditions (densities, temperatures, etc.) that influence the reference/ calibration.
The next time a calibration is performed, you will need to duplicate the conditions, or account for the
differences.
(b) All changes made to factory-default settings such as time constant, reference constants (loading factor),
etc.
(c) A record of "counts" being received from the detector may assist with future troubleshooting efforts.