3 Phase Power Analyser / EN
7 09/30/16 Version No. 00
Phase angle measurement (
φ
) :
Range
Resolution Calculated Accuracy
0.1
°
to +180
°
, 0.1
°
to -180
°
0.1
°
±15dgt
Measurement method : Calculated from power factor COSφ,
φ = COS
-1
PF.
Display item : Phase angle of each channel and its sum of multiple channels.
Polarity display :
For phase lag : No symbol
For phase lead : “-”.
Frequency measurement (Hz) :
Range
Resolution Accuracy
Measurement source
50Hz
0.1Hz ±0.1%rdg±2dgt
Voltage U1 > 50V
Measurable input range : > 50V
Three Phase Sequence Detection :
Input voltage range
Normal phase indication
Reverse phase indication
Measurement source
3P > 50V to 1000V
U1, U2 and U3
Active Power Energy measurement (kWh) :
Range
Resolution
Active power accuracy
Timer interval
Timer Accuracy
0.001 to 9.999KWh
0.001KWh
0.01 to 99.99KWh
0.01KWh
0.1 to 999.9KWh
0.1KWh
0.001 to 9.999MWh
0.001MWh
0.01 to 30.59MWh
0.01MWh
±1.0%rdg±20dgts
(10 to 299.9A)
±1.5%rdg±20dgts
(300 to 999.9A)
1 sec ±50ppm (25
°
C)
Measurement display : Display all active power consumption energy (sum of absolute values).
Apparent Power Energy measurement (kVAh) :
Range
Resolution
Apparent power
accuracy
Timer interval
Timer Accuracy
0.001 to 9.999kVAh
0.001kVAh
0.01 to 99.99kVAh
0.01kVAh
0.1 to 999.9kVAh
0.1kVAh
0.001 to 9.999MVAh
0.001MVAh
0.01 to 30.59MVAh
0.01MVAh
±1.0%rdg±20dgts
(10 to 299.9A)
±1.5%rdg±20dgts
(300 to 999.9A)
1 sec ±50ppm (25
°
C, 77
°
F)
Measurement display : Display all apparent power energy (sum of absolute values).