Piranha/Cobia & Quick Check Getting Started
Quick Check
20
The following meter settings are available in Quick Check:
Add a delay after the detection of trig before measurement of kVp starts. This will
delay the kVp measurement, it doesn't affect dose, mAs or time measurements.
This is the time the meter waits after trig off before it assumes that the exposure is
finished. The post delay must be set to a time longer than any dead time in the
radiation.
Current kV range. You need to change this only for radiography/fluoroscopy, for
mammography, dental and CT you only have one range.
Available calibrations for the internal (kVp and exposure) used. See table below that
describes usage of the different calibrations.
Sensitivity setting for the kVp detector (internal detector).
Hi = High sensitivity - for low dose rate
Lo = Low sensitivity - for high dose rate
Sensitivity setting for the dose and total filtration measurement (internal detector).
Hi = High sensitivity - for low dose rate
Lo = Low sensitivity - for high dose rate
Sensitivity setting for the external probe
Hi = High sensitivity - for low signals
Lo = Low sensitivity - for high signals
General (user-defined) correction factor used for all exposure related parameters
measured with the internal detector.
Beam Correction
factor
(External)
General (user-defined) correction factor used for all exposure related parameters
measured with the external detector.