22
Code
Calibration
Usage
C3
Straton (Siem1)
Suitable for Siemens CT with Straton tube
C4
GECT (7°)
Suitable for GE CT tubes with a 7° anode angle as well as for other
manufactures CT tubes and replacement tubes with a 7° anode angle
C5
Aquillion 64-
Suitable for Toshiba Aquillion 64-320 CT
C6
GECT (10.5°)
Suitable for GE CT tubes with a 10.5° anode angle
C7
M1
Mo/30 µm Mo
General mammography
M3
Mo/25 µm Rh
General mammography
M4
Rh/25 µm Rh
General mammography
M6
W/50 µm Rh
General mammography - suitable for Hologic Selenia Dimensions and
Fujifilm Amulet
M7
W/0.50 mm Al
General mammography - suitable for Philips MicroDose (Sectra)
M8
Mo/1 mm Al
General mammography
M10
W/50 µm Ag
General mammography - suitable for Hologic Selenia Dimensions and
Fujifilm Amulet
M11
W/75 µm Ag
General mammography
M12
W/50 µm Rh (Gio)
Suitable for Giotto Mammography
M15
W/0.70 mm Al
General mammography - suitable for Hologic Selenia Dimensions and
Fujifilm Amulet
M16
W/50 µm Ag (Sel)
Suitable for Hologic Selenia
M17
W/50 µm Rh (Sel)
Suitable for Hologic Selenia
M18
W/0.30 mm Cu
General mammography - Suitable for Hologic Selenia Dimensions and
Fujifilm Amulet
M19
W/0.70 mm Al (Inno.)
Suitable for Fujifilm Innovality
M20
W/50 µm Ag (Inno.)
Suitable for Fujifilm Innovality
M21
Mo/25 µm Rh (Sel)
Suitable for Hologic Selenia
M22
Rh/Ag (GE HC)
Suitable for GE Senographe Prestina
M23
Rh/Ag IQST (GE HC)
Suitable for GE Senographe Prestina
M24
Mo/Cu (GE HC)
Suitable for GE Senographe Prestina
M25
Rh/Cu (GE HC)
Suitable for GE Senographe Prestina
M26
Mo/Mo (GE HC)
Suitable for GE Senographe Prestina
M27
Affirm Prone W/Ag
Suitable for Hologic Affirm Prone
M28
Affirm Prone W/Al
Suitable for Hologic Affirm Prone
There are three different measuring modes available using the Piranha. They are as follows:
Measuring mode
Description and use
Normal
The Normal mode is used for short and long (fluoro) exposures. In this mode, your
meter will automatically sense if there is a signal and when it is above a certain trigger
level. If the exposure is long, the displays/grid will be updated with new data every 2
seconds. If the exposure is short, the results are displayed as soon as the trigger is
off.