CUSTOMIZATION OF SETTINGS
3.12 Revised 3/30/2010
Setting the Dispense Point
To select the initial position at which to start
dispensing mounting medium onto the glass slide, use this
function. The setting should be changed under certain
conditions such as when air bubbles are present on the
front end of the side of the slide or mounting medium
seeps out, especially at the front end of the slide (Figure
3-AA).
An optimum point can be set for each size of cover
glass.
NOTE: This menu item is available only when the system
is on standby or paused or when the system check screen
is displayed.
If air bubbles enter on the front side of the slide,
move the point toward the front end of the slide.
If mounting medium seeps out on the front side
of the slide, move the point towards the end of
the slide.
1. Press the MENU key twice.
2. Press the < and > keys to highlight 4, “Dispense
point, and then press the ENTER key.
3. The dispense start point setting screen appears
(Figure 3-BB).
4. Pressing the < key moves the dispense point towards
the front end of the slide, while pressing the > key
moves the dispense point towards the back end of the
slide.
Figure 3-AA
5. Press the ENTER key to save the selected setting
(Figure 3-CC). To return to the menu screen without
saving the change, press the EXIT key.
Figure 3-BB: Dispense Point Setting Screen
Figure 3-CC: Dispense Point Selection Screen
<Dispense point>60mm
[START] to start
<Dispense point>60mm
005
[<][>] to select
[ENTER] to save
First dispense

Other manuals for Sakura Tissue-Tek Glas g2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Glas g2 and is the answer not in the manual?

Sakura Tissue-Tek Glas g2 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Glas g2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals