5-20
PCU-110 Operation Manual
5) *OPC? Operation Complete Query
Syntax *OPC?
Description Places 1 in the output queue when all operation processing
has completed.
Response 0: (In operation)
1: (Operation completed)
6) *CLS Clear Status
Syntax *CLS
Description Clears all event registers and queues (except for the output
queue) and reects the summary in the Status Byte Register.
Clears the following items.
-Status Byte Register (except MAV)
-Standard Event Status Register
-Error Queue
Parameter None
Response None
7) *ESE Standard Event Enable Register Setting
Syntax *ESE<wsp><value>
Description Sets the Standard Event Enable Register (SEER).
Parameter Setting value from 0 to 255
Response None
8) *ESE? Standard Event Enable Register Query
Syntax *ESE?
Description Places the value of the Standard Event Enable Register
(SEER) in the output queue.
Parameter None
Response Integer from 0 to 255
9) *ESR? Standard Event Status Register Query
Syntax *ESR?
Description Places the value of the Standard Event Status Register (
SESR) in the output queue. Register is cleared after being
read.
Parameter None
Response Integer from 0 to 255