EasyManua.ls Logo

Santec PCU-110 - Page 41

Default Icon
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Polarization Control Unit PCU-110
5-23
5-6-5
PCU-110 Specic Command – Detailed Description
Specic commands are described in detail as follows.
1) Optical output related commands
:POLarization
Syntax [:OUTput]:POLarization<wsp><value>
Description Sets polarization state.
Parameter 1: Vertical Linear Polarization
2: Horizontal Linear Polarization
3: +45°Linear Polarization
4: -45°Linear Polarization
5: Right Hand Circular Polarization
6: Left Hand Circular Polarization
Response None
Example :POL 1
Sets the polarization state to Vertical Linear Polarization.
:POLarization?
Syntax [:OUTput]:POLarization?
Description Reads out polarization state
Parameter None
Response 1: Vertical Linear Polarization
2: Horizontal Linear Polarization
3: +45°Linear Polarization
4: -45°Linear Polarization
5: Right Hand Circular Polarization
6: Left Hand Circular Polarization
Example :POL?
+1
The polarization state is Vertical Linear Polarization.
:POWer:UNIT
Syntax [:SENSe]:POWer:UNIT <wsp><value>
Description Sets the power unit.
Parameter 0: dBm
1: mW
Response None
Example :POW:UNIT 0
Sets the power unit to “dBm”.