- 104 -
O2 density recovery data
0
5
10
15
20
25
0 10203040
TimeMinutes
O2 density
%
1 small door 60sec. open
1 small door 30sec. open
A
T:20䉝䚷SV:37
CO2 setting: 5%(003MPa)
O2 setting: 5%(0.05MPa)
No load
Days before CO2 tank becomes empty
0
100
200
300
400
500
600
700
800
012345678910
Door open & shut per day (Frequency)
Days
A
T:20䉝䚷SV:37
CO2 setting: 5%
Gas leak: 4.7% per 30min.
No load
Days before N2/O2 tank becomes empty
0
2
4
6
8
10
12
14
16
18
20
012345678910
Door open & shut per day (Frequency)
Days
A
T:20䉝䚷SV:37
CO2 setting: 5%
O2 setting: 5%
Gas leak: 4.8% per 30min.
No load