- 88 -
How to install H
2
O
2
decontamination kit
MCO-HL
and Electric lock
㻌㻌㻌
䐟 Mount transformer on plate meeting and
secure with screws (M4) and connect
each wires (red and brown) to the places
as left picture specified.
䐠 Remove rear cover.
䐡 Remove Inner cabinet rear cover, 2pcs o
frame under cover and door under cover.
Fit water proof connector to the place as
the arrow pointed in the picture.
Note) Mark triangle should be oriented to
upper position.
Fit 2 caps (M4) to the holes at cover
support.
Cover support
Cap (M4)
Frame under cover
Inner cabinet rear cover
Water proo
connector
Door under cover
䞉㻌 Philips Screwdriver
䞉㻌 Wrench (8mm)
Connector (White, 3P)
Wire (Red)
Wire (Brown)
Screw (M4)
Rear cover
Plate meeting Transformer Lock and key
Water proof
connector
Name plate
Caution label
Cap (M4)
Screw
M5
Screw (M4)
䕔Parts enclosures 䕔Tools