- 3 -
ŶControl specifications
Model MCO-19M MCO-19M(UV)
Thermal sensor Thermistor
Temperature display LCD digital display
CO
2
controller Microcomputer; Input by keypad
CO
2
density setting range: 0
䠂䡚
20
䠂䠄
Resolution; 0.1
䠂䠅
Non-volatile memory
CO
2
density display LCD digital display (Resolution; 0.1
䠂
)
Water level detection Water level sensor
䠄
Optic type
䠅
H
2
O
2
decontamination
(Option)
- Microwaved H
2
O
2
decontamination
- Low density H
2
O
2
solution (Oxidor) which can be purchased in the market
- Solution level detection by sensor
Electric lock (Option) Outer door is locked during H
2
O
2
decontamination
UV system Option. UV lamp lighting time can be
extended automatically
UV lamp lighting time can be extended
automatically
Self-diagnosis - Error code and error information are displayed by text.
- Remote alarm contact activates (DC30V, 2A)
㻌
with buzzer sounds.
ŶPerformance specifications
Model MCO-19M MCO-19M(UV)
Temperature control range Ambient temp. +5
䡚
50
o
C
䠄
ambient temp; 5
䡚
35
o
C
䠅
Temperature distribution ±0.25
o
C
䠄
ambient temp; 25
o
C
䚸
setting; 37
o
C, CO
2
5%, O
2
5%, no load
䠅
Temperature variation ±0.1
o
C
䠄
ambient temp; 25
o
C
䚸
setting; 37
o
C, CO
2
5%, O
2
5%, no load
䠅
CO
2
control range 0
䡚
20 %
CO
2
variation ±0.15 %
䠄
ambient temp; 25
o
C, setting; 37
o
C, CO
2
5%, O
2
5%, no load
䠅
O
2
control range 1~18, 22~80 %
O
2
variation ±0.2 %
䠄
ambient temp; 25
o
C, setting; 37
o
C, CO
2
5%, O
2
5%, no load
䠅
Chamber humidity 95±5%R.H.
Rated voltage, frequency 110~120V, 60Hz 220V, 60Hz 220~240V, 50Hz
Amperage Max. 3.1A Max. 1.5A Max. 1.5A
Heat emission Max. 1274 kJ/h Max. 1206 kJ/h
Power consumption Max. 354 W Max. 335W
Noise level 30 dB (A scale)
Usable environment condition
Temperature: 5~35
o
C, Humidity: Equal or less than 80%R.H.
䠄
Designed performance may not be obtained when ambient temp. is equal
or lower than 15
o
C
䠅
Note) Specifications will be subject to change without notice.