- 8 -
O
2
span calibration
(ExWhen setting O
2
density is 5.0% and actual O
2
density is 4.5%:
Process Operation MENU
1 Put densitometer thorough the sample port at the front side of the unit to measure
chamber O
2
density.
2
Select Tools in the Top Screen.
Tools
3
Select Tools (1/2) is displayed. Select temp / CO
2
Calibration and press OK key.
OK
4 By pressing digit shift key and numerical value shift key, input actual O
2
density
4.5% in the column O
2
Span. Select OK and press ENTER key.
OK
5
Automatically reverts to Select Tools (1/2) screen.
*Note (1) Input range: +1.0% ~ -1.0% (per input).
(2) Repeat O2 calibration for several times until displayed CO2 density will match with
actual density.
(3) It is unacceptable to commence O2 span calibration when setting O2 density deviates
rom 18.0 ~ 22.0.
㻌㻌㻌There is a blank in the rignt side of O2 Span column.
Display of Process-2
Display of Process-3
O
2
span calibration