Sanyo MCO-19M - Setting Procedures for Optional Components

Sanyo MCO-19M
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
- 77 -
Setting procedures for optional components
Applicable for; MCO-19UVS, MCO-HL, MCO-HP, MCO-SG
1. How to access to optional components
(1) Input of service code
Operation Key Display
1 ----- “Top Screen”
2 Press MENU key. MENU “MENU” window
3 Select “Tool” and press ENTER key. ENTER Select “Tool” screen
4 Press “MENU” key. MENU “MENU” window
5 Select “SVC” and press ENTER key. Password input screen
6 Input service code ‘384’ by numeric keys. Numeric key
7 Press MENU key. MENU “MENU” window
8 Select OK and press ENTER key. ENTER Service mode “Select Tools Svc”
Note) If you fails to input service code, “Wrong Password” is displayed under the “Password” on the
screen and error tone emits. The password “384” should be input correctly.
Service code is unchangeable.
(2) Access to optional components
# Operation Key Display
1 ----- Service mode “Select Tools Svc”
2 Select “System Configuration” and press
ENTER key.
ENTER “System configuration” window
3 Select applied optional components* from
“System Configuration” and input “1” by
numeric key.
Numeric key “MENU” window
4 Press MENU key. MENU “MENU” window
5 Select “OK” and press ENTER key. ENTER
* Optional components in “System Configuration”
Optional components Application in “System Configuration
UV lamp add-on kit (MCO-19UVS) “UV Function”
H
2
O
2
generator (MCO-HP),
H
2
O
2
Decontamination kit (MCO-HL)
“Decon Function
Automatic CO
2
Density Calibration Add-on Kit (MCO-SG) “Std Gas Cal Function”
Automatic CO
2
Cylinder Changeover System (MCO-21GC) “CO2 Gas AutoChanger”