- 78 -
2. Operation check for optional components
(1) UV lamp add-on kit (MCO-19UVS)
zUV setting
# Operation Key Display
1 ----- “Top Screen”
2 Press MENU key. MENU “MENU” window
3 Select Tool and press ENTER key. ENTER “Select Tool” screen
4 Select “UV Setting” and press ENTER key. ENTER “UV setting” screen
5 Select “UV Timer” and input setting period
by numeric keys. Press ENTER key.
Numeric keys
ENTER
6 Press MENU key. MENU “MENU” window
7 Select “OK” and press ENTER key. ENTER
Note 1) UV lamp lighting time range is 0~30 minutes. Factory default is 5 minutes.
If you set lighting time to ‘0’, UV lamp will not illuminate.
2) UV lamp switches off by opening the outer door during UV lamp illuminates.
UV lamp illuminates for predetermined time after the outer door is closed.
3) If UV lamp lighting time is set longer than 5 minutes, or if the outer is open and close
frequently, it may be caused there is condensation in the chamber, or temperature
distribution is ununiformly, or UV lamp lifetime to be shorted by heat emission from the UV
lamp.
4) Consult a SANYO representative or agent for replacement of UV lamp.
5) UV lamp occasionally stays illuminating longer than the predetermined time for the
compensation of reducing UV illumination which is caused by increasing UV lamp
accumulation time.
zOperation check
# Operation
1 Press door switch when the outer door is open and the inner door is closed.
2 After several seconds pass, check if ‘UV indicator’ is displayed on LCD display.
3 You can see blue ray at the front of humidifying tray cover.
Never look UV ray directly.