- 79 -
(2) H
2
O
2
generator (MCO-HP), H
2
O
2
Decontamination kit (MCO-HL)
zInstallation
Fill 80cc distilled water in MCO-HP. By using the cable enclosed, connect MCO-HP with the connector
on the bottom right of the incubator. (See picture as below)
Connector cap should be kept in safe place.
Ensure 2 pins of MCO-HP should be aligned to the holes on the rear duct. If not so, it may be caused
poor in spray performance.
For the check of spray performance, hold door switch by the tape.
zOperation check
# Operation Key Display/Remark
1 ----- “Top Screen”
2 Press H
2
O
2
key for 3 seconds. H2O2 “H2O2 Decontamination”
(System check starts)
3 H2O2 Decontamination screen
‘H2O2 Decon’ blinks
----- (If error message is shown, refer
to Instruction Manual P.57.)
4 Ensure if the outer door is locked by
electric key.
-----
5 Open outer door by using key to check
spray performance is correctly obtained.
----- Turn key toward the direction
when you locked and take key off
before door shut. You cannot turn
the key with the door shut.
H
2
O
2
generator
UV lamp
Connector
Ultrasound oscillator
Front
Front
H
2
O
2
generator
Rear duct
UV lamp