- 87 -
Installation of H
2
O
2
Generator (MCO-HP)
Prior to use H
2
O
2
Generator, ensure to check the following notes.
After installation
࣭Make sure the condensing fan should be correctly installed to the position.
࣭Make sure the duct should be correctly mounted to the position.
࣭Make sure the H
2
O
2
Generator should be fixed by 2 fixtures which are pointed in the picture and
each fixture should be correctly mounted to the holes at the bottom of the duct.
“Err34” (The volume of H
2
O
2
generated is abnormal)
࣭ If you find an error message, ‘Err34’ after predetermined decontamination process is
complete, the voltage of the apparatus may be out of the specification and it is necessary to
adjust H
2
O
2
generating time. Please consult to your authorized sales representative.
Duct
H
2
O
2
Generator (MCO-HP)
Fixture㸦left and right㸧
* Secure H
2
O
2
Generator with the fixtures to prevent generator from being moved or rattled.