- 87 -
Installation of H
2
O
2
Generator (MCO-HP)
Prior to use H
2
O
2
Generator, ensure to check the following notes.
After installation
Make sure the condensing fan should be correctly installed to the position.
࣭Make sure the duct should be correctly mounted to the position.
Make sure the H
2
O
2
Generator should be fixed by 2 fixtures which are pointed in the picture and
each fixture should be correctly mounted to the holes at the bottom of the duct.
“Err34” (The volume of H
2
O
2
generated is abnormal)
࣭ If you find an error message, ‘Err34’ after predetermined decontamination process is
complete, the voltage of the apparatus may be out of the specification and it is necessary to
adjust H
2
O
2
generating time. Please consult to your authorized sales representative.
Duct
H
2
O
2
Generator (MCO-HP)
Fixtureleft and right
* Secure H
2
O
2
Generator with the fixtures to prevent generator from being moved or rattled.