<Setting of chamber temperature>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting temperature in the right side of “Temperature “ column
in “Stand-by Setting” display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 0 ~ 50.0䉝.
Note) Maximum 50.0䉝 for settable value. Error tone is not emit if you input a value over 50.0䉝.
<Setting of CO2 density>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting CO2 density in the right side of “CO2 Density” column
In “Stand-by Setting” display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 0 ~ 20.0%.
Note) Maximum 20.0% for settable value. Error tone is not emit if you input a value over 20.0%.
<Setting of O2 density>
In Top Screen, press MENU key and select Set.
Press ENTER key to display current setting O2 density in the right side of “O2 Density” column
In “Stand-by Setting” display.
Input required value and press MENU key and select OK in reverse video.
Press ENTER key to memorize the input value.
Setting range is 1~80.0%. (Controllable range: 1~18.0%, 22.0%~80.0%)
Note) O2 is set to 1.0% when you input a value lower than 1.0% and is set to 80.0% when you input a value
over 80.0%. Error tone is not emit.
Note) Press ENTER key with “Cancel” displayed in reverse video to cancel the setting change.
<Display of high limit setting temperature>
High limit setting temperature is displayed in the right side of “High Limit” column in “Stand-by Setting” display.
Setting range is 35䉝~51䉝.
When a chamber temperature is over high limit setting temperature, LED illuminates and audible alarm sounds
continuously. (The audible alarm will not stop sounding if BUZZER key is pressed)