㻌 㻌䠘Transferring 1day Log data to PC>
Note) It is necessary to set Hyper terminal in PC side.
In Top Screen is displayed and press MENU key and select “Log”.
Press ENTER key to display Log data.
Display required Log data by pressing left and right shift key (䊹㻌 䊻).
Display MENU key and select “PC 1D” and press ENTER key to display “Progress” in the upper left side.
Select “Start” in MENU key and press ENTER key to transfer 1day Log data to PC with displaying “Sending..”
and quantity of transferring data in the lower side.
“Finished..” will be displayed when transferring 1day Log data is complete.
Note) Select “Cancel” in MENU key and press ENTER key to return to Top Screen.
㻌
䢴 䢶䢴䢳䢲
䣥 䣧 䣮䣅 䣣 䣰
䣅 䣮 䣧 䣣 䣴
䣃 䣮 䣮䣒 䣅
䣒 䣅 䢳 䣆
䣗䣏 䣇 䣐
䢳 䢹䢲 䢱 䢷 䢱䢴 䢲 䢳䣅䢷 䢲 䤑
㻌
<Transferring all Log data to PC>
Note) It is necessary to set Hyper terminal in PC side.
In Top Screen is displayed and press MENU key and select “Log”.
Press ENTER key to display Log data.
Display required Log data by pressing left and right shift key (䊹㻌 䊻).
Display MENU key and select “PC All” and press ENTER key to display “Progress” in the upper left side.
Select “Start” in MENU key and press ENTER key to transfer all Log data to PC with displaying “Sending..”
and quantity of transferring data in the lower side.
“Finished..” will be displayed when transferring 1day Log data is complete.
Note) Select “Cancel” in MENU key and press ENTER key to return to Top Screen.
<Setting in PC for transferring log data> (For Windows 2000 and Windows XP)
From Window Start Button, select Program – Accessories – Communications – HiperTerminal to start
the HyperTerminal. (If the HyperTerminal is not available from the Start Menu, execute the following file:
C:¥Program Files¥Windows NT¥hypertrm.exe.)
In the HyperTerminal Window, set a new connection, the name (for example, Sanyo), connection settings,
Method of connection, COM1, properties of COM1, and port.
Baud rate: 9,600, Data bits: 8, Parity:None, Stop bits:1, Flow control: Xon/Xoff
(The communications condition of the Incubator will be set automatically to the above settings when the
Progress Display appears.)
Note) Optional RS232C/RS485 Interface MTR-480 (RS-232C/RS-485) is required to transfer data.