Ensure to install an optional Automatic CO
2
density Calibration Add-on Kit䠄MCO-SG䠅 when standard gas
calibration is done automatically.㻌
Note) “Standard Gas Calibration” display will not be appeared and a device will automatically return to Top Screen
when you select “1” in “Std Gas Cal Function” column in the “System Configuration” display.
㻌
㻌
Operational sequence for standard gas calibration
Step 0 1 2 3 4 5
Status
Normal
operation
Pressure check
in standard gas
cylinder
Stand-by for
Start key
pressed
Injecting
standard gas
CO2/O2
calibration
(Pressure check
for standard gas
cylinder)
Standard gas
calibration is
complete
Standard
gas control
valve
OFF
ĺĺĺ
ĺĺĺ
ON
ĺĺĺ
OFF
Display
“Std Gas Calib
System
Check…”
“Std Gas
Calib Start?”
“Std Gas
Calibrating…”
“Std Gas
Calibrating…”
“Std Gas Calib
Finished.”
Note
“Err41: Std Gas
Empty” is
appeared when
a cylinder
becomes empty
“Err41: Std Gas
Empty” is
appeared when a
cylinder becomes
empty
㻌
Error message
Error ʋ Status Display Action
Err41
When standard gas cylinder becomes empty
(When a gas pressure is lower than standard)
Gas pressure in the cylinder is checked by
pressure switch.
”Std Gas Calib Error”
”Err41:StdGas Empty” is
appeared on the message
bar.
Automatic calibration
is inactive