H2O2 decontamination
Procedures MENU key
0 Locate trays and humidifying pan to the position.
Fill 6% H2O2 solution in optional H2O2 unit (MCO-HP).
Connect the cable to the port in the inner rear wall and close both inner and outer door.
1 Press H2O2 key for 3 seconds in Top Screen.
2 “H2O2 Decon System Check…” is displayed and an appratus ensures to connect the cable
correctly or to check H2O2 solution level or to check door open.
3 “Ready to Start?” is displayed and press MENU key to select “OK” and press ENTER key. OK
4 Warming up and H2O2 misting is commenced.
“H2O2 Decon” blinks in reverse video in the upper left corner of the display.
5 An apparatus stops H2O2 misting when it reaches predetermined amount and starts
UV resolution. “UV Resolve” blinks in reverse video in the upper center of the display.
6 “Please Remove H2O2 Unit” is displayed after H2O2 decontamination is complete.
Open both outer and inner door, then disconnect the cable from H2O2 unit.
7 “PUSH MENU OK” is displayed in the lower left side of the display.
Press MENU key and select “OK”.
OK
8 Press ENTER key to return to Top Screen.
Step 0 1 2 3 4 5 6 7 8
Status
Run in the
specificati
on
Check
H2O2 unit
Stand-by Warming up
Primary
H2O2
misting
Primary
H2O2
preservati
on
Secondary
H2O2
misting
UV
resolution
Finish
Ultrasonic
oscillator
OFF ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ON OFF ON OFF OFF
UV
Lamp
OFF ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ON OFF
Chamber
temp.
37.0䉝
ĺ ĺ ĺ ĺ ĺ ĺ
37.0䉝
Ļ
45.0䉝
45.0䉝
ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ
45.0䉝
Electric
lock
OFF ĺ ĺ ĺ ĺ ĺ ĺ ON ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ ĺ OFF
Display
Warming up
temp. blinks
“H2O2
Decon”
blinks n
reverse
video
“Decon”
blinks in
reverse
video
“H2O2
Decon”
blinks n
reverse
video
“UV
Resolve”
Blinks in
reverse
video
Diagnosis
to proceed
next step
Select
“Start OK”
“Decon”
temperature
is +/-1.0䉝
H2O2
resolution
level or
H2O2
misting time
is over
15min.
(Skip) (Skip)
Note) In default setting, Step 5 (Primary H2O2 preservation time) and 6 (Secondary H2O2 misting time) are set in 0
minute, so it means that Step 7 will be commenced after Step 4.
When both setting times for Step 5 and 6 are 0 minute, the display will be changed from “Step 5” to “Step 6” if
primary H2O2 misting is not complete until predetermined time expires.
Set temperature will return to 37䉝 when press MENU key and select “OK” and press ENTER key in Step 8.