12
INCUBATOR COMPONENTS
Switch box
1. Power switch (POWER)
The switch for ON/OFF of all power source including the plug outlet.
2. Remote alarm terminals
Alarm signal can be drawn out via the contact output. Permissible contact capacity : DC30 V・2 A
a) Contact output: Connect the lead wires to COM and NO terminals.
(open when normal condition, close when abnormal condition)
b) Contact output: Connect the lead wires to COM and NC terminals.
(close when normal condition, open when abnormal condition)
COM and NO terminals are closed at the failure of electricity supply.
3. Glow starter (type: FG-1P)
1
2 3