SETTING OF KEY LOCK (MENU/Std-by)
Caution: The buzzer rings for a long time when a wrong password is input. Input a correct password.
The password for key unlock must be shared and administered by all users on this product.
The setting of key unlock when shipped from the factory is “0000”.
Refer to page 40 for changing the password.
PROCEDURES OF SETTING VALUE
Setting of chamber temperature
Procedures of changing chamber temperature from 25 to -10 are shown below. As for other values,
the same procedure is applied.
Procedures of changing chamber temperature from 25 to -10
Operation Key LCD Display
1 ---- Top screen is displayed.
2 Press the menu button (MENU). MENU MENU window is displayed.
3
Select Std-by in MENU window and
press the enter key (ENTER).
ENTER
Stand-by Setting screen is displayed and chamber
temperature setting value is changed into white
letters.
4
Change the value from 25.0 to 10.0
pressing the character input key.
Chamber temperature setting value 10.0 is
displayed.
5 Press the menu button (MENU). MENU MENU window is displayed.
6
Select / in MENU window and
press the enter key(ENTER).
ENTER
Chamber temperature setting value -10.0 is
displayed.
7
Press the enter keyENTER.
ENTER
Next itemLight Stepis displayed in white letters.
8 Press the menu button (MENU). MENU MENU window is displayed.
9
Select OK in MENU window and press
the enter key (ENTER).
ENTER
The set value is memorized and top screen is
displayed.
CAUTION
The settable range of chamber temperature is between -10 and +60
.
The lighting is available only with temperature range between +2 and +50
o
C when the optional light
add-on kit (MIR-L15) is installed. In the case of temperature out of range, the lighting is not usable.
The lamp is off with the temperature out of range when the saved program is running.
In this case, temperature fluctuation may exceed
±
1.5
o
C or it may take longer time to pull down the
chamber temperature.
The temperature control range is -10+50
o
C when the optional inner door (MIR-55ID) is installed.
The temperature is limited to 50
o
C when the saved program includes temperature setting over 50
o
C. In
this case, the temperature setting over 50
o
C is indicated by blinking.