VARIOUS SETTING (MENU/Tools)
Setting of date, time, log (Tools/Date Time)
1. Select “Date Time” in the Select Tools screen. Press the menu button (MENU) to show the menu
window. Select “OK”, and press the enter key (ENTER).
2. The Date Time screen is displayed. Set date, time or log cycle.
Date input (Ex: Feb. 14, 2010)
Input 100214 in the date cell.
Time input (Ex: 12:15:00)
Input 121500 in the time cell.
Log cycle input (Ex: 6 minutes)
Input 6 in log Interval cell.
Note:
The default is 6 minutes.
The acceptable range is between 2 and 30 minutes.
Relation between the log interval and spans that can be memorized
1: Log interval 2 min About 5 days
2: Log interval 6 min About 14 days
3: Log interval 30 min About 70 days
After passing the memory limit, the older data is deleted and newer data is memorized.
Setting of automatic defrost
Select one of two defrosting patterns. The default is 0(Manual defrost). For the details of automatic
defrost, refer to page 3334.
0: Manual defrost
1: Automatic defrost
S e l e c t T o o l s MIR-554
L o g
D a t e T i m e
A l a r m ,
O p t . , L o w H u mi . S e t t i n g
K e y L o c k P
S e t ting
D e f a u l t S e t t i n g
D e l e t e U s e r D a t a
ME NU
O K
Svc
Ca nc e l
D a t e T im e
D a t e
1 0 / 0 1 / 0 1 ( YY/MM/DD)
T i m e 1 2 : 0 0 : 0 0 ( h h : m m : s s )
L o g I n te r v a l 6min (2-30min)
D e f T i me r 1 (0.Manual 1.Auto)
3:00