START-UP
OF
UNIT
Follow
the
procedures
for
the
initial
and
consequent
operations
of
the
unit.
1.
C
onnect
the
power
cord
to
the
dedicated
outlet
with
appropriate
rating.
Acaution
If
the
unit
is
unplugged
or
the
power
to
the
unit
is
interrupted,
do
not
restart
the
unit
for
at
least
5
minutes.
This
protects
the
compressor.
2.
On
start-up,
the
alarm
buzzer
sometimes
operates.
In
this
case,
stop
the
buzzer
by
pressing
the
alarm
buzzer
stop
key
(BUZZER).
3.
Set
the
chamber
temperature
to
5°C.
4.
Allow
the
chamber
temperature
to
fall
to
5°C.
Check
the
chamber
temperature
on
the
temperature
indicator.
5.
Turn
on
the
fluorescent
light
switch
to
check
the
light.
After
checking,
turn
off
the
switch
if
the
light
is
not
necessary.
6.
Set
the
desired
temperature.
When
the
chamber
temperature
gets
to
the
set
temperature,
begin
slowly
placing
items
into
the
chamber
to
minimize
the
temperature
rise.
^CAUTION
The
chamber
is
refrigerated
by
the
forced
circulation
of
cooled
air
inside
the
chamber.
Ensure
that
the
air
intake
and
exhaust
vents
are
not
blocked.
Adequate
space
should
be
provided
between
the
items
inside
the
unit
to
allow
air
circulation.
If
any
electrical
equipment
is
used
in
the
chamber,
the
chamber
temperature
cannot
be
maintained
due
to
the
heat
emission
from
the
equipment.
Refer
to
the
table
below
when
you
use
any
equipment.
1
(at
ambient
temperature
of
35°C)
Heat
load
in
chamber
Lowest
chamber
temperature
100
W
2°C
200
W
9°C
300
W
18°C
52
14