CHAMBER
TEMPERATURE
SETTING
Table
1
shows
the
basic
procedure
for
setting
the
chamber
temperature.
Perform
key
operations
in
the
sequence
indicated
in
the
table.
The
example
in
the
table
is
based
on
the
assumption
that
the
desired
temperature
is
4°C.
Note:
The
unit
is
set
at
the
factory
with
a
chamber
temperature
of
5°C.
Table
1
Basic
operation
sequence
(Example:
Chamber
set
temperature
4°C)
Description
of
operation
Key
operated
Indication
after
operation
1
Turn
the
power
switch
ON.
The
current
chamber
temperature
is
displayed.
1
2
0
2
Press
set
key.
SET
The
first
digit
is
flashed.
m
m
a
"
v
Press
numerical
value
shift
key
and
When
pressed,
the
figure
of
settable
digit
increases.
3
1
0
0
4
scroll
the
figure
to
4.
Set
temperature
is
memorized
and
the
current
chamber
temperature
is
4
Press
set
key.
SET
2
0
displayed.
Note:
The
temperature
set
mode
returns
to
the
temperature
display
mode
automatically
when
90
seconds
has
passed
without
any
key
operation.
In
this
case,
the
chamber
temperature
setting
is
not
accepted.
The
available
set
range
of
chamber
temperature
is
between
2
and
23°C.
The
partial
freezing
may
occurred
if
the
set
temperature
is
lower
than
3°C.
KEY
LOCK
FUNCTION
This
unit
is
provided
with
a
key
lock
function.
When
the
key
lock
is
ON,
change
of
temperature
setting
Display
Mode
Function
L0
Key
lock
is
OFF
Enable
to
chanqe
of
temperature
setting
LI
Key
lock
is
ON
Disable
to
chanqe
of
temperature
setting
Table
2
Procedure
for
key
lock
setting
(change
from
key
lock
OFF
to
key
lock
ON)
Description
of
operation
Key
operated
Indication
after
operation
1
The
current
chamber
temperature
is
displayed.
mm
_
Ej
2
Press
digit
shift
key
for
5
seconds.
VP1
The
first
digit
is
flashed.
g
MSB!
3
Press
numerical
value
shift
key
and
scroll
the
fiqure
to
1.
1
When
pressed,
the
figure
of
settable
digit
increases.
i
B
The
key
lock
is
set
to
ON.
4
Press
set
key.
SET
The
current
chamber
temperature
is
H
displayed.
_
_
L
54