Sanyo MPR-721 - SETTING OF RINGBACK OF ALARM BUZZER; OPERATION CHECK AFTER RECOVERY

Sanyo MPR-721
39 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SETTING
OF
RINGBACK
OF
ALARM
BUZZER
The
alarm
buzzer
operates
again
after
certain
period
(ringback
time)
even
if
the
alarm
buzzer
is
silenced
by
pressing
the
alarm
buzzer
stop
key
(BUZZER)
when
the
same
alarm
status
is
continued.
Set
the
ringback
time
to
prevent
the
misidentify
the
alarm
status.
Setting
range
of
ringback
time:
between
10
and
60
minutes
(10
minutes
interval)
Display
of
setting:
between
010
and
060
(000
display
shows
no
ringback)
Initial
setting
(factory
setting):
30
minutes
The
alarm
buzzer
will
not
recover
once
the
alarm
buzzer
is
silenced
by
pressing
the
alarm
buzzer
stop
key
(BUZZER)
when
the
ringback
time
is
set
to
000.
However,
the
alarm
buzzer
will
operate
if
other
alarm
status
is
detected.
Table
6
Procedure
for
setting
of
ringback
of
alarm
buzzer
(Change
the
ringback
time
to
20
min.
from
30
min.)
___
Description
of
operation
Indication
after
operation
1
The
current
chamber
temperature
is
displayed.
w
2
Press
up
arrow
key
for
5
seconds.
t
F00
is
displayed
and
the
first
digit
blinks.
F00
3
Press
up
arrow
key
5
times.
t
The
display
is
changed
to
F05
from
F00.
si/
F0S
/|V
D
Press
scroll
key
once.
The
second
digit
blinks.
si/
F
05
/is
5
Press
up
arrow
key
twice.
1
The
display
is
changed
to
F25
from
F05.
si/
F0
5
7TK
6
Press
set
key.
SET
The
current
setting
(030)
is
displayed
and
the
second
digit
blinks.
sl/
030
/|V
B
Press
up
arrow
key
and
scroll
the
figure
to
2.
i
The
display
is
changed
to
020
from
030.
\l/
000
8
Press
set
key.
SET
Ringback
time
is
memorized
and
the
current
chamber
temperature
is
displayed.
Note:
The
setting
mode
returns
to
the
temperature
display
mode
automatically
when
90
seconds
has
passed
without
any
key
operation
(auto-return
function).
In
this
case,
the
setting
is
not
accepted.
The
setting
should
be
performed
during
normal
operation,
not
during
alarm
status.
OPERATION
CHECK
AFTER
RECOVERY
After
recovery
from
a
power
failure,
the
unit
will
resume
operation
automatically
with
the
setting
before
power
failure.
Accordingly,
there
is
no
need
for
re-set
however,
always
check
the
running
status
after
recovery.
The
set
value
is
memorized
by
nonvolatile
memory
during
power
failure.
19
-
57