ALARMS
&
SAFETY
FUNCTIONS
This
unit
has
the
alarms
and
safety
functions
shown
in
Table
7,
and
also
self
diagnostic
functions.
Table
7
Alarms
and
safety
functions
______
Alarm
&
Safety
Situation
Indication
Buzzer
Safety
operation
High
temperature
alarm
If
the
chamber
temperature
deviates
from
the
set
temperature
+2°C
or
up
to
+14°C.
Alarm
lamp
is
flashed.
Chamber
temperature
is
flashed.
Intermittent
tone
with
15
minutes
delay.
Remote
alarm
with
15
minutes
delay.
Low
temperature
alarm
If
the
chamber
temperature
deviates
from
the
set
temperature
-2°C
or
up
to
-14°C.
Alarm
lamp
is
flashed.
Intermittent
tone
with
15
minutes
delay.
Remote
alarm
with
15
minutes
delay.
Chamber
temperature
is
flashed.
Intermittent
tone
Remote
alarm
Over-heat
protection
When
the
chamber
temp,
is
higher
than
30°C.
(Reset
when
the
chamber
temp,
is
lower
than
25°C)
Inside
fan,
drain
pan
heater,
defrost
heater,
drain
pipe
heater
OFF
Over-cooling
protection
Compressor
OFF
Power
failure
alarm
When
the
power
to
the
unit
is
disconnected.
Remote
alarm.
Power
failure
alarm
(with
MPR-48B)
At
power
failure.
If
the
power
supply
cord
is
unplugged.
Alarm
lamp
blinks.
Intermittent
tone
Alarm
status.
Door
alarm
When
the
door
is
open.
Door
check
lamp
is
lit.
Intermittent
tone
with
2
minutes
delay.
Auto-return
When
there
is
no
key
pressing
in
each
setting
mode
for
90
seconds.
Chamber
temperature
is
displayed.
Finishing
of
each
setting
mode.
Key
lock
When
the
key
lock
is
"ON".
Change
of
setting
is
disable.
Sensor
abnormality
If
the
thermal
sensor
is
disconnected.
Alarm
lamp
is
flashed.
E01
and
50°C
(or
-50°C)
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
Unit
keeps
continuous
running.
If
the
thermal
sensor
is
short-circuited.
Alarm
lamp
is
flashed.
E02
and
50“C
(or
-50°C)
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
Unit
keeps
continuous
running.
If
the
defrost
sensor
is
disconnected.
Alarm
lamp
is
flashed.
E03
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
If
the
defrost
sensor
is
short
circuited.
Alarm
lamp
is
flashed.
E04
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
If
the
compressor
sensor
is
disconnected.
Alarm
lamp
is
flashed.
E05
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
If
the
compressor
sensor
is
short-circuited.
Alarm
lamp
is
flashed.
E06
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
If
the
ambient
temp,
sensor
is
disconnected.
Alarm
lamp
is
flashed.
E07
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
If
the
ambient
temp,
sensor
is
short-circuited.
Alarm
lamp
is
flashed.
E08
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
Memory
back-up
When
power
failure
or
disconnected
the
power
cord.
Set
value
memorized
Setting
check
(with
MPR-48B)
When
the
battery
switch
is
turned
ON
without
setting
after
attachment.
E09
and
chamber
temp,
is
displayed
alternately.
Condenser
temp,
abnormality
In
the
event
of
failure
of
fan
motor
for
cooling
the
compressor
E10
and
chamber
temp,
are
displayed
alternately.
Intermittent
tone
Remote
alarm.
Compressor
of
high
stage
side
stops.
Note:
The
remote
alarm
is
silenced
by
pressing
the
alarm
buzzer
stop
key
(BUZZER)
as
the
remote
alarm
is
operated
in
conjunction
with
the
buzzer,
except
for
the
power
failure
alarm.
When
more
than
two
alarm
conditions
occur
simultaneously,
the
lowest
number
error
code
has
priority
on
the
error
display.
After
a
power
failure,
the
unit
will
resume
operation
with
the
set
value
that
was
in
place
before
power
failure
occurred.
The
over-cooling
protection
may
activate
when
the
set
temperature
is
2°C.
In
this
case,
the
chamber
temperature
will
rise
to
6°C,
but
this
is
not
malfunction.
21
-
59