Additional functions: 
 
Beep Tone (Acoustic Signal) 
 
If you desire, you can turn off the beep tone – also called an acoustic signal. 
 
Acoustic signal  Code  Changes 
ON  8  2  1   * 
OFF  8  2  2   
 
 
Universal Switch 
 
An external switch can be connected via the interface port – you can change the setting in the menu code to 
alter the function of this switch 
 
Funktionen Code Changes 
Print  8  4  1   * 
Tare  8  4  2   
 
 
 
Power-On Mode 
 
Depending on the your operating requirements, you can change the power-on mode of your balance. 
 
lf you define the code for "toggle between on and off,"  the balance power will remain turned off after a power 
failure has occurred or after you have temporarily disconnected your balance from AC current. After the bal-
ance is turned off, power is not available for the "standby" mode. Use the ON/OFF key 
(16)
 to turn the bal-
ance on or off. 
 
In the setting "toggle between on and standby", the balance power is turned on automatically after a power 
failure has occurred or after you have disconnected your balance temporarily from AC power. 
 
In the setting "automatic power-on," the balance will automatically turn on again after a power failure has 
occurred, the balance has been disconnected from AC power, or after you have pressed the ON/OFF key 
(16)
. In this setting, the balance can no longer be turned off by the ON/OFF key. 
 
Power-on mode  Code  Changes 
(Power) off Æ on ÅÆ standby  8  6  1   * 
on ÅÆ (Power) off  8  6  2   
on ÅÆ standby  8  6  3   
Automatic power-on  8  6  4   
 
 
Automatic Shutoff 
(functions only if the menu code 8 6 2 is selected or if the balance is powered by the 
rechargeable battery pack) 
 
lf you operate the balance with a battery pack, set the menu code for the "Automatic Shutoff" function in or-
der to save energy and thereby increase the service life of the battery. lf the weight shown in the display 
does not change for 2 minutes, the display will show " 
 ". After another 2 minutes, the balance turns itself 
off automatically. 
 
Automatic Shutoff  Code  Changes 
ON  8  7  1   
OFF  8  7  2   * 
 
 
 
 
 