17
– Factory: set the pipette to the default factory settings, press the operating
button or OK.
– It is possible to store three (3) different adjustments in the adjustment
memory: ADJ1, ADJ2, ADJ3
– Activate desired memory place by using the adjustment wheel and pressing
the operating button.
– The pipette asks to set the custom adjustment values. Press the operating
button or OK to activate the values.
– Editing the adjustment values:
– Use the adjustment wheel to select the desired memory place (ADJ1,
ADJ2 or ADJ3) and press EDIT.
– Select the desired calibration (1-point, 2-points or 3-points)
– First display shows the target volume, press OK.
– Display shows the actual volume. Enter the volume by turning
adjustment wheel and press operating button or OK.
– Pipette will ask:
One (1) volume in 1-point adjustment
Two (2) volumes in 2-point adjustment
Three (3) volumes in 3-point adjustment
3.3.10.2 Sound
The sounds can be adjusted for the adjustment wheel and messages (errors
and notifications). It is also possible to turn all the sounds OFF.
– Select menu
– Select Setup
– Select sound
– Use the adjustment wheel to set the sounds ON/OFF.
– Press the operating button or OK to save the settings
– Press BACK - the settings are not changed
3.3.10.3 Backlight
Display colours equal to the colour coding of the display cover and operating
button colours. Colours are selected to indicate the tip compatibility.
Colour coding is active during charging, while pipetting the colour coding is
off and the display is white to maximise the contrast.
– Select menu
– Select Setup
– Select backlight
– Use the adjustment wheel to set the backlight.
If the display colour is changed, the display is always set to the selected colour
and the colour will not change to optimal contrast while pipetting.