10
3.3 Programming
3.3.1 Pipetting (forward pipetting)
Pipetting (p) aspirates the selected volume of liquid and dispenses it with the
blow out function. Recommended for aqueous liquids, liquids containing small
amounts of detergent or proteins and solvents.
Select Pipetting in main menu. The Pipetting mode is activated with the
latest settings.
Start pipetting by pressing the operating button.
Editing the mode:
Press EDIT button (middle soft key) or turn the adjustment wheel to the
end. Parameter background colour is highlighted to indicate that the
editing mode has been activated.
Turn the adjustment wheel to set the volume. Confirm the setting by
pressing the operating button or OK.
Press NEXT (middle soft key) to move the cursor to the next parameter to be
edited.
Press BACK to return to the previous screen without saving the changes.
Confirm the setting by pressing the operating button or OK.
Additional functions
Counter, Mixing and Tracker can be used in conjunction with the Pipetting
mode.
Press ADV in the right soft key to activate advanced functions.
Turn the adjustment wheel to move the cursor to the desired function and
confirm the selection by pressing the operating button or OK.
Counter counts the number of dispensings. Counting can be selected to
start in any desired number.
Mixing mixes the liquids manually or automatically. the mixing volume can
be adjusted according to the pipette’s maximum volume.
Tracker advises user in microplate dispensing by displaying the location for
the next dispensing.
NOTE! Only one additional feature can be selected at a time.
3.3.2 Reverse Pipetting
Aspirates the selected volume as well as an excess volume. Recommended for
biological, foaming and viscous liquids. In dispensing mode the excess is left
inside the tip and is then discarded.
Select Reverse mode in main menu. The Reverse mode is activated with the
latest settings.
Start pipetting by pressing the operating button.
Editing the mode:
Press EDIT button (the middle soft key) or turn the adjustment wheel to
the end. Parameter background colour is highlighted to indicate that
edit mode has been activated.
Turn the adjustment wheel to set the volume. Confirm the setting by
pressing the operating button or OK.
Press NEXT (middle soft key) to move the cursor to the next parameter to be
edited.
Press BACK to return to the previous screen without saving the changes.
Confirm the setting by pressing the operating button or OK.
Press MENU
Change settings and press OK or NEXT
Press EDIT
Choose mode